1. Remove the sample holder cover. |
2. Place the sample cuvette in the sample holder and adjust its distance from the lens using the Zaxis knob. |
3. Replace the cover. |
4. Turn the laser safety key to the On position to activate the laser. |
5. Set parameters manually or directly load any predefined paramter set. |
6. Start one-time scan or continous scan. |
7. Review the acquired Raman spectrums. |
8. Upload the data, or do a real-time analysis. |
Operating temperature range | -10 °C ~ 35 °C | Power Consumption | 25 W |
Spectral Range | 220-2200 cm-1 | Spectral Resolution | 2 cm-1 |
Grating | 1200 lines/mm | Spectral Line-width | < 0.2 nm |
CCD Pixel Array | 512 x 122 | Pixel Size | 24 x 24 microns |
Full Well Capacity | 300,000 electrons | quantum efficiency | 85% at 250 nm |
Dark Noise | 200 e-/pixel/sec at 0 °C | ADC resolution | 16 bit |
Lasing Wavelength | 785 nm | Power Output | 70 mW |
Integration Time | 50ms ~ 100s |